Description
PhysioGo.Lite Laser is the newest low-level laser therapy device in ASTAR's offer. Its unique design and safety concepts ensure patient safety during therapy. The compact size of the laser makes it an ideal device for physiotherapy centres and for home environmental therapy.
Benefits for the patient
Laser applicators - great therapeutic possibilities
Clinical studies and many years of experience in the use of laser therapy have demonstrated the effectiveness of this therapy for the appropriate wavelengths. The so-called "therapeutic window" has been defined. This is the wavelength range from 650 nm to 1200 nm. The PhysioGo.Lite laser, thanks to its cooperation with all applicators available in the ASTAR offer: point probes, scanning laser applicator, cluster laser applicator, stimulates tissues with a wavelength of 660 nm (R) - superficial effect and 808 nm (IR) therapy for deep tissues. In addition to the wavelength, the power of the applicator is also an important parameter. The therapist working with the PhysioGo.Lite laser in combination with the CL-1800 (cluster laser applicator) has a power of up to 1800 mW.
In addition, the use of the scanning laser applicator makes it possible to treat large surfaces with simultaneous use of radio and infrared radiation. Three different laser sources can be connected to the PhysioGo.Lite Laser, which are available from ASTAR.
Safety first - a self-check procedure
The device software allows you to perform a self-test every time you switch it on. This is an ongoing monitoring of the device's performance, which makes the PhysioGo.Lite Laser a safe tool in the hands of the physiotherapist. After the self-test is complete, the display shows information on the readiness of the accessories connected to the device. The device automatically checks the laser power before starting treatment procedures. In addition, the user can measure the power of the point sensor at any time, which is an additional functional advantage.
Medical devices require a periodic annual inspection. To meet these requirements, PhysioGo.Lite Laser allows you to save the date of the next inspection for your device. When the date has passed, the device will automatically remind the user to carry out the required inspection. Like all ASTAR products, the device complies with the requirements of the Medical Device Directive.
Benefits for the therapist
Home treatment
In healthcare facilities that provide services to patients with limited mobility or at home, the PhysioGo. Lite Laser will work perfectly thanks to its size, light weight and special PRO bag that contains the device and accessories. In addition, it is possible to extend the functionality of the device with a rechargeable battery module. This allows the device to operate for up to 4 hours without being connected to a power source.
Modern and ergonomic
The PhysioGo.Lite Laser is equipped with a 5-inch colour display and touch panel. The laser can be operated in manual or software mode. Innovative changes in the software make the device's operation very intuitive. Moreover, the device automatically calculates the duration of treatment depending on: power, dose, frequency and area of the therapeutic field. At any time, the user can check the statistics of the procedures performed and the operating time of individual laser sources of low laser therapy.
Standby mode
The PhysioGo.Lite Laser is equipped with a rechargeable battery module and has the ability to enable battery saving mode. After selecting the appropriate option in the service menu, the device will enter the STANDBY mode after a specified period of laser inactivity.
Special mode for optical fibre applicators
In the case of point probes, the use of fibre-optic applicators significantly increases the therapeutic capabilities of the device. PhysioGo.Lite Laser has a special mode implemented in the device software that guarantees the accuracy of the parameters during therapy. Thanks to this, the device can be successfully used for laser puncture or laser therapy in ENT.
Medical encyclopedia - defined therapeutic procedures
There are 175 built-in treatment programmes, illustrated by a treatment encyclopedia, to optimally adapt the parameters of low-level laser therapy to the diseases of the patients being treated. The therapist can select a specific therapy from the device's memory by entering the name of the relevant disease or by dividing it into areas such as orthopaedics, sports medicine, aesthetic medicine, rheumatology, neurology, dermatology and angiology. The user can create and store up to 200 custom treatment programmes in the PhysioGo.Lite Laser unit's memory. The device also has predefined laser programmes - 8 Nogier acupuncture programmes and 30 Foll acupuncture programmes.
- Compatible with applicators: scanning, shower and spot probes
- Radiation mode: continuous and pulsed
- Adjusting the laser power
- Adjustment of the filling
- Automatic check of laser power
- Automatic calculation of time according to treatment parameters - dose, power, filling, treatment area
- Three modes of influence on the treatment area in scanning applicators
- Dedicated modes for interaction with fibre optic applicators
- Fibre optic handpieces for laser puncture and ENT applications
- Pilot beam indicating the place of application
Detailed technical specifications
independent treatment channels 1
manual mode
Laser therapy
Work with applicators: scanning laser, cluster laser and spot probes
Preset treatment programmes
built-in treatment programmes, including: 175
Programmes for measuring IR points 39
Programmes for measuring the R-point are 18
programmes with a Nogier frequency 8
applications with the Voll frequency 30
Cluster laser applicator applications 54
sequence of programmes for scanning laser applicators 26
user-configurable programmes - 200
favourite apps
class of the laser device 3B
treatment timer 1 s - 100 minutes
Laser therapy parameters - biostimulation with laser point probes
wavelength of the red light laser dot probes 660 nm
maximum power output of red light sensors 80 mW
infrared laser point wavelength probes 808 nm
maximum power of infrared spot probes - 1.5 kW 400 mW
power control - - power control 25%, 50%, 75%, 100%
frequency of the pulse mode 1 - 5000 Hz
fill factor in pulse mode - 1.5 10 – 90%
Laser therapy parameters - scanning laser applicator
wavelength of the scanning laser applicator 808 and 660 nm
maximum power of the scanning laser applicator 450 and 100 mW
power control - - power control 50%, 100%
frequency of the pulse mode 1 - 5000 Hz
Filling factor for a scanning laser applicator in pulse mode 75%
Laser therapy parameters - cluster laser applicator
The wavelength of the cluster laser applicator is 4x 808 nm and 5x 660 nm
maximum power of the cluster laser applicator 4x 400 mW and 5x 40 mW
power control - - power control 50%, 100%
frequency of the pulse mode 1 - 5000 Hz
fill factor in pulse mode - 1.5 10 – 90%
General technical parameters
dimensions 25 x 27 x 16.5 cm
the weight of the device is 3 kg
battery type (optional) - Lithium-ion
battery capacity (optional) 2100 mAh
power supply, energy consumption 100-240 V, 50/60 Hz